Lens and sensibility: Compact high-efficiency illumination system for in-orbit microscopy

DOI: 10.22443/rms.mmc2025.297

Authors

Ewan Drever-Smith (1), Daniel J. Robson (2), Flavius C. Pascut (1), Chantal Cappelletti (3), Veeren M. Chauhan (2), Kevin F. Webb (1)

Affiliations

  1. Optics & Photonics Research Group, Faculty of Engineering, University of Nottingham
  2. School of Pharmacy, Faculty of Science, University of Nottingham
  3. Nottingham Geospatial Institute, Faculty of Engineering, University of Nottingham

Space travel poses a set of unique challenges. Due to microgravity, muscles and bones experience less mechanical stress. This can cause muscles to atrophy and bones to lose mineral density – increasing operational and post-flight risks to crew. To facilitate the continued human presence in low-earth orbit, as well as future missions to deep-space, the effects that the stressor-rich space environment have on biological systems must be better understood [1]. 

Caenorhabditis elegans is a model organism, about 1mm in length, that has been used for decades to study the muscular, genomic and neuro-physical changes of animals (including humans) in space and on Earth. The WormSail CubeSat Mission, a collaborative CubeSat project between the University of Nottingham and the University of Brasilia [2], aimed to develop a microsatellite capable of sustaining live C. elegans within its payload, enabling their behaviours and physiology to be studied in the deep-space environment. Here, we present the development of a proof-of-concept microscopy system to be implemented within the form factor of a 1U (10x10x10cm, <2kg) CubeSat in order to allow their longitudinal monitoring. 

This use case imposed several challenges that arose from the restrictive form factor, limited power access, and necessary environmental resilience. These constraints directly challenged the design aspects of a platform utilising conventional imaging techniques and thus demanded a radical redesign. Figure 1 presents the illuminator and optical design resulting from careful consideration of these constraints. A custom illuminator was created using blue (465 ±20nm) and green (520 ±30nm) through-hole LEDs (Nichia Corporation), housed within a custom 3D-printed housing. These LEDs were arranged within two distinct elements to enable multi-modal imaging: an outer annulus comprising 16 alternating green/blue LEDs and an inner region made from four quadrants of Spectralon diffuser material, each backlit by a green LED. These together comprised a multi-modal transmission illumination system capable of dark-field, bright-field, fluorescence, and differential phase contrast (DPC) illumination. 

Figure 1 – Optical setup and illuminator imaging modalities demonstration. A) Optical arrangement following a condenser-free schema. The central quadrants of the illuminator are diffuser segments, each back-lit by a single LED. The quadrants fall within the acceptance NA cone of the objective, whereas the ring of blue/green LEDs falls outside the pupil (dark-field configuration). B) Proof-of-concept implementation of the optical arrangement, measuring 46mm top to bottom. C) Photograph of the prototype illuminator. D) Images of an AmScope MR095 calibration slide containing a 100 x 10 μm division scale (red), yellow fluorescent highlighter ink cut-through by the edge of a glass coverslip (yellow) and a drop of yellow highlighter ink (blue), acquired using the prototype illuminator in each mode of operation (v-viii); the corresponding view of the illuminator in each mode is vertically orientated to each image (i-iv).

The illuminator’s outer ring of LEDs was used for condenser-free dark-field imaging [3]. Diascopic fluorescence imaging was achieved by replacing alternate LEDs in the dark-field ring with excitation LEDs for the fluorophore (e.g. GFP), in combination with an emission interference filter (510 ±20nm) between the objective and tube lenses. This dark-field diascopic fluorescence illumination (Figure 1Diii) allowed use of excitation filters to be foregone with minimal compromise, since direct light from the LEDs did not reach the camera sensor. Abrogation of the epi-illumination path further improved the system’s compactness.

Quantitative differential phase contrast (qDPC) imaging was achieved using the array of central diffusing panels, by acquiring images using each half-circle illumination pattern (Figure 1Div), processed using DPC algorithms [4] in order to recover a quantitative phase image (Figure 1Dviii). Each required half-circle pattern was formed using two adjacent diffuser quadrants. Since the objective back focal plane could not be directly visualised using this minimalist configuration, we used the dark-field illumination mode to position the illuminator in the z-dimension; ensuring that, when dark-field conditions were reached, the central quadrants completely filled the objective NA. This improved the accuracy of the phase recovery, especially at low spatial frequencies. Bright-field imaging was achieved by lighting all four panels in unison. The illumination LEDs for each mode were controlled by a micro-processor (Raspberry Pi 4B). 

Figure 2 – Multi-modal images of C. elegans, illustrating the illuminator’s optics-agnosticism. A) Multi-modal images of C. elegans specimens acquired using custom illuminator, Raspberry Pi Camera HQ and optical system of Olympus IX71 inverted microscope for 3 different objective lens magnifications. B) Z-alignment of the illuminator for 5X/0.13NA, 10X/0.3NA and 20X/0.4NA objective lenses.

Figure 1D (v-viii) demonstrates multi-modality, where each complementary imaging method reveals additional information on the specimen. The condenser-free operating mode also allowed flexible deployment of the prototype illuminator on a research-grade microscope (Olympus BX1: Figure 2), where bright-field, dark-field and qDPC images of C. elegans in a petri dish were acquired using 5x, 10x, and 20x magnification objective lenses. The only adjustment required was a realignment in the position of the illuminator in the z-direction.

In conclusion, a multi-modal imaging system was delivered with an overall optic axis of 46mm, enabling dark-field, bright-field, diascopic fluorescence, and quantitative phase (DPC) imaging with no moving parts. Thus, we have demonstrated a compact optical system utilising a custom illuminator that enables high-contrast, multi-modal imaging techniques whilst adhering to the form factor limitations imposed by a 1U CubeSat; the size of the system (including microprocessor) was 56x87x62mm with a total mass of 157g and peak power usage of 4.05W at 5V. This form factor could be further reduced, if necessary, by substituting the Raspberry Pi 4B microprocessor for a more compact alternative. The use case of the system could also be expanded to terrestrial applications, especially in limited-resource environments given the use of 3D-printed and commercial-off-the-shelf components.

References

[1]    NASA, ‘Moon to Mars Objectives’. Sep. 2022. Accessed: Feb. 11, 2025. [Online]. Available: https://www.nasa.gov/wp-content/uploads/2022/09/m2m-objectives-exec-summary.pdf

[2]    Robson, D., et al. "The WormSail CubeSat: An International Educational Project To Elevate Space Science And Education." 72nd International Astronautical Congress. 2021.

[3]    K. F. Webb, ‘Condenser‐free contrast methods for transmitted‐light microscopy’, Journal of Microscopy, vol. 257, no. 1, pp. 8–22, Jan. 2015, doi: 10.1111/jmi.12181.

[4]    L. Tian and L. Waller, ‘Quantitative differential phase contrast imaging in an LED array microscope’, Opt. Express, vol. 23, no. 9, p. 11394, May 2015, doi: 10.1364/OE.23.011394.


Keywords

Quantitative Phase Imaging

Multi-modal

3D printed